Measurement variables |
Dissolved oxygen, saturation, oxygen partial pressure |
Measuring range |
0 - 20mg/L,0 - 200%,0 - 400hPa |
Resolution |
0.01mg/L,0.1%,1hPa |
Accuracy |
±1.5%FS |
Repetition |
±0.5%FS |
Temperature measurement type |
NTC 10K/PT1000 |
Auto A/manual H Measuring range |
-10 - 60℃ Resolution: 0.1℃ Correction accuracy: ±0.5℃ |
Output type |
4 - 20mA current transmission output |
Max.loop resistance |
750Ω |
Accuracy |
0.1%FS |
Output type |
RS485 digital signal output |
Communication protocol |
Standard MODBUS-RTU (customizable) |
Power supply |
AC220V±10%,50Hz/60Hz (optional 24V±10%) |
Alarm relay |
AC250V、3A |
Wiring Method
①Dissolved oxygen transmitter
Open a 92.5 * 92.5(mm) installation hole on the instrument cabinet or installation panel
(the dimension is 100*100*150mm). Insert the instrument into the installation hole and
latch on the butterfly clasp, as shown below.